Bad Science: The instrumental data in the Floyd Landis case

Bad Science : The instrumental data in the Floyd Landis case / Robert D. Blackledge. – (Clinica Chimica Acta (2009) 406 (22 May) : p. 8-13)

  • doi: 10.1016/j.cca.2009.05.016. Epub 2009 May 22

Background: In 2006 Floyd Landis won the world's most prestigious bicycle race, the Tour de France. However, not many days after the race's conclusion it was released to the press that the Laboratoire National de Dépistage du Dopage (LNDD) had found Landis' urine after stage 17 positive for synthetic testosterone.
Methods: This review examines the instrumental data and methodology used by LNDD in the Landis case. The conclusions reached by LNDD were based on results of 2 separate instrumental methods. Subsequent to urine extraction and possibly derivatization, samples were initially screened via gas chromatography/mass spectrometry (GC/MS) using selected ion monitoring (SIM) to measure the ratio of testosterone to epitestosterone (T/E). Final confirmation of exogenous testosterone was determined by measuring the (13)C/(12)C stable isotope ratios in 4 metabolites of testosterone via gas chromatography combustion stable isotope ratio mass spectrometry (GC-C-IRMS).
Conclusion: T/E ratios determined by LNDD in Landis' stage 17 urine were unreliable due to the combined factors of an unsatisfactory extraction, high GC background, failure to obtain baseline peak separation for epitestosterone, unreliable quantization of the epitestosterone peak due to both peak overlap and because it was barely above background noise, and because only a single ion mass (432) rather than a minimum of 3 was used for SIM (in violation of both LNDD's SOP and WADA procedures). GC-C-IRMS methodology is less well known to the analytical chemistry community, but here too the results obtained by LNDD were unreliable. GC-C-IRMS errors can be briefly summarized as uncertain peak identification, unsuitable standards, and unreliable (and possibly biased) calculation of (13)C/(12)C ratios due to peak overlap as well as LNDD's usage of manual peak integration rather than use of the instrument system software.

Comment in:

Advocacy versus impartial scientific review: a problem for scientists and the courts / Larry D. Bowers. – (Clinica Chimica Acta (2009) 406 (19 May) ; p. 14-17)

  • doi: 10.1016/j.cca.2009.05.006. Epub 2009 May 19

Floyd Landis: an unsafe conviction, regardless of the quality of the data / N.K. Faber. – (Clinica Chimica Acta (2010) 411 (12 October) ; p. 117-118)

  • doi: 10.1016/j.cca.2009.10.001. Epub 2009 Oct 12.

Parameters

Science
Case report
Date
22 May 2009
People
Blackledge, Robert D.
Country
United States of America
Language
English
ADRV
Adverse Analytical Finding / presence
Sport/IFs
Cycling (UCI) - International Cycling Union
Laboratories
Paris, France: Agence Française de Lutte contre le Dopage (AFLD)
Analytical aspects
Mass spectrometry analysis
Reliability of the testing method / testing result
Doping classes
S1. Anabolic Agents
Substances
T/E ratio (testosterone / epitestosterone)
Testosterone
Document category
Abstract
Document type
Pdf file
Date generated
7 May 2014
Date of last modification
7 September 2020
Category
  • Legal Source
  • Education
  • Science
  • Statistics
  • History
Country & language
  • Country
  • Language
Other filters
  • ADRV
  • Legal Terms
  • Sport/IFs
  • Other organisations
  • Laboratories
  • Analytical aspects
  • Doping classes
  • Substances
  • Medical terms
  • Various
  • Version
  • Document category
  • Document type
Publication period
Origin